DB-ALC1 / DB-ALC2

Definition

Examples of manufacturer-specific blood-alcohol chromatography columns. Different stationary-phase selectivities can help reveal interference. Verify proprietary chemistry and method suitability from current documentation.

Manufacturer
Agilent
DB-ALC1 phase
Polyethylene glycol
DB-ALC2 phase
Nitroterephthalic acid-modified PEG
Use
Dual-column confirmation of blood ethanol

Common questions

Why does blood-alcohol testing use two columns with different phases instead of confirming on the same column twice?+

Running the sample on two columns with different stationary-phase chemistry gives orthogonal selectivity, so a co-eluting interferent that shares ethanol's retention time on one column is unlikely to coincide on the other, which lets analysts confirm the ethanol peak identity independent of any single-column interference.

What result on DB-ALC1 and DB-ALC2 would raise concern rather than confirm an ethanol finding?+

If the retention times on the two columns produce inconsistent quantitation, for example a peak identified as ethanol on one column but not corroborated at a matching concentration on the other, that discrepancy signals a possible co-eluting interferent and the result would need further investigation before being reported as confirmed.

Related terms

2100:1 Partition Ratio
A conventional blood-to-breath concentration ratio, not a universal physiological constant. At this ratio, 2100 volumes of breath contain the same ethanol mass...
Beta Elimination Rate (Beta)
The rate at which BAC falls during the post-absorptive phase of alcohol metabolism, expressed as mg/100 mL/hour. Population mean approximately 15 mg/100...
Blood Alcohol Concentration (BAC)
Mass of ethanol per unit volume of whole blood, expressed as mg per 100 mL (equivalent to mg/dL or mg%), or as...
Internal Standard (IS)
A compound chemically near-identical to the analyte, often a deuterated isotopologue (d3-morphine, d5-EDDP, d5-alprazolam, pyrene-d10), spiked into every sample at a known...
NIST SRM 1828
A series of NIST-certified reference material aqueous ethanol standards (0.02 to 0.40 per cent w/v) used to calibrate forensic blood ethanol analytical...
Partition Coefficient (K)
Here defined as equilibrium liquid concentration divided by gas concentration. It depends on analyte, matrix and conditions; state the convention before interpretation.
Sodium Fluoride / Potassium Oxalate (Grey-Top Tube)
The standard blood collection tube for forensic ethanol analysis. Sodium fluoride inhibits microbial fermentation (which would produce artefactual ethanol post-collection); potassium oxalate...
Static Headspace
A sample preparation technique in which a sealed vial is heated until volatile analytes reach thermodynamic equilibrium between the liquid matrix and...