Rayleigh Criterion

Definition

The condition that two point sources are just resolved when the central maximum of one Airy disc coincides with the first minimum of the other, giving a minimum resolvable distance r = 0.61 lambda / NA.

Formula
r = 0.61 lambda / NA
Defines
Minimum resolvable distance between two points
Condition
First Airy minimum meets adjacent central maximum
Depends on
Wavelength and numerical aperture

Common questions

How does the Rayleigh criterion affect the choice of microscope objective in trace evidence work?+

Because resolution improves with higher numerical aperture and shorter wavelength, an examiner needing to distinguish fine structural detail in a fibre or particle selects a higher-NA objective or shorter-wavelength illumination to push the minimum resolvable distance smaller.

What happens to two point sources closer together than the Rayleigh limit?+

Their diffraction patterns overlap enough that they blur into what appears to be a single feature, so detail below that limit cannot be resolved as two separate points regardless of magnification.

Related terms

Abbe Diffraction Limit
The minimum period of a periodic grating structure that can be resolved: d = lambda / (2 NA). Sets the hard optical...
ASTM E766
The ASTM International standard practice for image-scale calibration in microscopy, extended by OSAC to govern optical microscope calibration in US forensic trace-evidence...
Depth of Field
Axial thickness of the specimen that appears acceptably in focus at one time. High in stereoscopic microscopes, very low in oil-immersion compound...
Empty Magnification
Magnification beyond approximately 1000 times the objective NA; enlarges the image without adding resolvable structural information, producing blurred detail that can be...
Infinity-Corrected Optics
An objective design in which the back focal plane produces a collimated (afocal) beam rather than a converging beam, allowing optical elements...
ISO 8576
The ISO standard governing the accuracy of stage micrometers and reticle eyepieces, specifying Class I and Class II tolerance levels traceable to...
Koehler Illumination
The standard illumination method for research microscopy in which the light source is focused at the condenser back focal plane, producing uniform...
Numerical Aperture (NA)
The product of the refractive index of the medium between objective and specimen and the sine of the half-angle of the maximum...
Parfocality
The property of an objective nosepiece turret such that switching objectives keeps the specimen approximately in focus without re-focusing; defined by ISO...
Plan Objective
An objective corrected for field curvature so that the entire image field is simultaneously in focus; required for photomicrographic evidence documentation to...
Stage Micrometer
A calibrated glass slide with a ruled scale of known spacing (typically 0.01 mm divisions), used to establish the pixels-per-micrometre calibration factor...