GRIM-3

Definition

Glass Refractive Index Measurement, third-generation instrument by Foster + Freeman (UK); an automated phase-contrast hot-stage system that identifies the RI match point by curve-fitting the phase-contrast edge signal vs temperature, achieving ±0.00015 RI unit precision.

Manufacturer
Foster + Freeman Ltd (UK)
Measurement method
Automated phase-contrast microscopy with heated immersion-oil stage
Typical precision
0.0001 to 0.00015 RI units

Common questions

What does GRIM-3 measure?+

GRIM-3 measures the refractive index of glass fragments using an automated hot-stage system. It identifies the exact RI value where a glass sample has the same light-bending properties as the surrounding immersion oil, which is the match point for forensic comparison.

How precise is GRIM-3?+

GRIM-3 achieves precision of about 0.0001 to 0.00015 refractive index units. This high precision allows forensic labs to distinguish between glass samples from different sources.

How does GRIM-3 work?+

The instrument uses automated phase-contrast microscopy with a heated immersion-oil stage. As temperature changes, the phase-contrast edge in the microscope shifts. GRIM-3 automatically curve-fits this signal against temperature to find the exact RI match point.

Related terms

ASTM E1967
The ASTM International standard for forensic glass RI measurement; specifies three acceptable methods (automated, semi-automated, manual), minimum precision of ±0.0001 RI units,...
ENFSI ENG3-2013
The ENFSI Glass Expert Working Group guideline for glass comparison casework; specifies the examination sequence, within-source variation requirements, statistical comparison framework, and...
ASTM E2927
Standard test method for the determination of trace elements in glass samples using LA-ICP-MS; specifies the element suite, calibration standards (NIST SRM...
Bayesian Likelihood Ratio
A statistical framework for expressing the evidential value of a comparison: the ratio of the probability of the observed RI similarity given...
Bayesian Likelihood Ratio (LR)
The statistical framework that evaluates glass-fragment evidence by comparing the probability of the observed fragment population assuming contact with the source glass...
Becke Line
The bright optical halo at the edge of a solid fragment immersed in a liquid, visible under the microscope; moves toward the...
Cauchy Equation
The empirical equation n(λ) = A + B/λ² + C/λ⁴ that describes the wavelength dependence (dispersion) of a material's refractive index; used...
Dispersion
The variation of refractive index with wavelength; all glasses and immersion oils show dispersion, which is why RI must be specified at...
Hotelling's T2
The multivariate generalisation of the t-test used to compare mean element-concentration vectors between questioned and known glass populations; accounts for the full...
Immersion Method
The forensic technique of matching a glass fragment's RI to that of a reference oil by adjusting temperature; the fragment becomes invisible...
LA-ICP-MS
Laser ablation inductively coupled plasma mass spectrometry; a technique that fires a laser pulse at a sample surface, vaporises a tiny amount,...
Locke 1985 Protocol
The calibrated immersion-oil series and operational procedure for forensic glass RI measurement described by Locke and Underhill (1985); specifies the oil series,...