Absorbance (A)

Definition

The base-10 logarithm of the ratio of incident to transmitted light intensity, A = log10(I0/I). Dimensionless. Linear with concentration in the working range; departs from linearity above roughly A = 1.5 because of stray light, detector saturation and analyte self-absorption.

Related terms

Bandpass and Slit Width
The wavelength interval the monochromator delivers to the sample. Routine forensic UV-Vis runs at 1 nm or 2 nm bandpass; pharmacopoeial methods...
Chromophore
The molecular subunit responsible for a dye's colour, typically a conjugated double-bond system; chemical erasure destroys the chromophore while leaving the decolourised...
Diode Array Detector (DAD)
A UV-Vis detector that records the full 190 to 800 nm absorbance spectrum of every peak as it elutes, instead of a...
Molar Absorptivity (ε)
The proportionality constant in the Beer-Lambert law, in L mol⁻¹ cm⁻¹. A property of the analyte at a given wavelength, solvent and...
Photomultiplier Tube (PMT)
Cascaded electron-multiplication detector behind the monochromator slit on a dispersive scanning instrument. Reads one wavelength at a time at very high sensitivity.