MRM (Multiple Reaction Monitoring)

Definition

A targeted MS/MS scan mode on a triple quadrupole that monitors specific precursor-to-product ion transitions, with two transitions per analyte (quantifier and qualifier) plus a deuterated internal standard treated as the BSA 2023 Section 63 minimum for confirmatory toxicology reports in India.

Instrument
Triple quadrupole mass spectrometer
Transitions per analyte
Minimum two: quantifier and qualifier
Confirmation requires
Two transitions plus retention-time match
Common use
Opioid and toxicology analysis

Common questions

How is MRM different from a regular MS scan?+

MRM filters for specific ion pairs on a triple quadrupole instrument. The first filter (Q1) catches a chosen precursor ion, the collision cell breaks it apart, and the third filter (Q3) catches a specific product ion. This targeting gives much higher sensitivity and specificity than scanning all ions.

Why do forensic labs use two transitions per substance in MRM?+

One transition is the quantifier (measures how much substance is present), and one is the qualifier (confirms you have the right substance). Combined with a retention-time match, two transitions are the standard requirement to confirm identification and rule out false positives.

What's an MRM transition exactly?+

It is one precursor-to-product ion pair. A transition tells the instrument to select ions of mass X in Q1, fragment them in the collision cell, then select ions of mass Y in Q3. Each substance gets at least two transitions running at the same time.

Related terms

4-Anilinopiperidine Scaffold
The structural core conserved across fentanyl and its analogues: a piperidine ring bearing an anilide group at the 4-position. Clandestine variation at...
6-Monoacetylmorphine (6-MAM)
A transient diagnostic metabolite of heroin (diacetylmorphine) formed by the rapid hydrolysis of one acetyl group. Half-life is 6 to 25 minutes...
Accurate Mass and Mass Defect
On a Q-TOF or Orbitrap, the measured m/z is reported to four decimal places at parts-per-million accuracy. The deviation from the nominal...
Alkaloid Ratio Profiling
A source-attribution technique using the morphine:codeine:papaverine:noscapine ratio in heroin seizures to infer the geographic origin of the opium feedstock, reflecting the chemotype...
Atmospheric Pressure Chemical Ionisation (APCI)
A complementary atmospheric-pressure source that vaporises the LC eluent and uses a corona discharge to ionise a reagent gas, which transfers charge...
ATR (Attenuated Total Reflectance)
An FTIR sampling mode in which the sample is pressed against a diamond or ZnSe crystal and the evanescent wave probes the...
Beer-Lambert Law
A = εbc. Absorbance equals molar absorptivity times path length times concentration. The linear relation between absorbance and concentration that makes UV-Vis...
Carfentanil
A fentanyl analogue with a 4-carbomethoxy substituent on the piperidine ring; approximately 10,000 times the analgesic potency of morphine. Implicated in the...
CID and HCD
Collision-induced dissociation. The precursor is accelerated and collides with a neutral gas (nitrogen, argon, helium) in the collision cell. Internal energy converts...
Deuterated Internal Standard
An isotopologue of the analyte where two to five hydrogens are replaced by deuterium. Same chromatographic and ionisation behaviour, slightly higher mass....
Diacetylmorphine
The IUPAC name for heroin; the 3,6-diacetyl ester of morphine, produced by acetylation with acetic anhydride. More lipophilic than morphine, crosses the...
Dose Heterogeneity
The variation in active ingredient content between nominally identical tablets in a counterfeit or clandestinely pressed batch. High dose heterogeneity in illicit...

Explained in these topics

Your journey to becoming a forensic professional starts here.

Practice with mock tests, learn from structured notes, and get your questions answered by a global forensic community, all in one place.